Evaluation of a new blood-conserving arterial line system for patients in intensive care units
- 1 April 1993
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 21 (4) , 507-511
- https://doi.org/10.1097/00003246-199304000-00008
Abstract
To evaluate blood samples obtained from a new blood-conserving arterial line system for the presence of hemodilution or heparin contamination. Prospective, clinical trial. A coronary intensive care unit in a tertiary-care teaching hospital. Cardiovascular patients in whom invasive arterial blood pressure monitoring was indicated. Paired blood samples were obtained from a conventional arterial line system and a new blood-conserving arterial line system for the measurement of hematocrit and partial thromboplastin time, and compared to evaluate for the presence of either hemodilution or heparin contamination. A Bland-Altman bias analysis of the variability between the two blood draw methods was performed. The analysis indicated that a) a randomly determined partial thromboplastin time obtained from the blood-conserving arterial line would lie between 3.32 and −5.11 of the partial thromboplastin time taken from the conventional arterial line value with 95% confidence; and b) a randomly determined hematocrit obtained from the blood-conserving arterial line would lie between 1.97 and −1.85 of the hematocrit taken from the conventional arterial line value with 95% confidence. We conclude that a) blood samples obtained with the blood-conserving arterial line demonstrate no evidence of hemodilution or heparin contamination; b) the blood-conserving arterial line provides blood samples without the need for an initial volume of blood to be discarded; c) the blood-conserving arterial line provides a means for blood conservation in the intensive care setting. (Crit Care Med 1993; 21:507–511)Keywords
This publication has 0 references indexed in Scilit: