Automated non-invasive measurement of cardiac output by the carbon dioxide rebreathing method: comparisons with dye dilution and thermodilution.
Open Access
- 1 March 1990
- Vol. 63 (3) , 195-199
- https://doi.org/10.1136/hrt.63.3.195
Abstract
The accuracy and reproducibility of indirect measurement of cardiac output at rest by the carbon dioxide rebreathing (indirect Fick) method with an automated respiratory analysis system (Gould 9000IV) were compared with simultaneous measurements made in duplicate by dye dilution and thermodilution in 25 patients having cardiac catheterisation studies. Measurements of cardiac output by the carbon dioxide rebreathing method were not significantly different from those obtained with dye dilution (mean difference -0.3 l/min, SD 0.76, 95% confidence interval -0.7 to 0.1). Thermodilution significantly over-estimated cardiac output by a mean of 2.2 l/min or 39% (SD 1.5, 95% confidence interval 1.6 to 2.8) compared with the carbon dioxide rebreathing method and significantly overestimated cardiac output by 1.9 l/min or 31% (SD 1.2, 95% confidence interval 1.2 to 2.5) compared with dye dilution. The reproducibility of measurements of cardiac output in individual patients was satisfactory with the dye dilution method but was poor with carbon dioxide rebreathing and thermodilution. Indirect measurement of resting cardiac output by the Gould 9000IV automated carbon dioxide rebreathing method is more accurate but the variability inherent with this method requires that multiple measurements be taken for each determination. Measurement of cardiac output by the thermodilution method by a commercially available cardiac output computer was not satisfactory because not only was there considerable variability between repeat measurements but the method also consistently overestimated cardiac output compared with the dye dilution method.This publication has 32 references indexed in Scilit:
- Application of noninvasive techniques for measuring cardiac output in hypertensive patientsAmerican Heart Journal, 1988
- Doppler Estimation of Cardiac Output: Principles and PitfallsEchocardiography, 1987
- Comparison of direct fick and thermodilution cardiac output techniques at high flow ratesThe American Journal of Cardiology, 1987
- Limitations of cardiac output measurements by thermodilutionCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1986
- Validation of a compact system for measuring gas exchangeCritical Care Medicine, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- The reproducibility of nongeometric analysis of cardiac output and left ventricular volume by radionuclide angiographyAmerican Heart Journal, 1985
- Doppler echocardiographic measurement of cardiac output: Comparison with a non-golden standardThe American Journal of Cardiology, 1984
- Statistics and ethics in medical research: V--Analysing data.BMJ, 1980
- Relative importance of nervous control of cardiac output and arterial pressureThe American Journal of Cardiology, 1971