Effect of half-wavelength membranes on the axial resolution of real-time ultrasonic scanners
- 1 September 1983
- journal article
- research article
- Published by Acoustical Society of America (ASA) in The Journal of the Acoustical Society of America
- Vol. 74 (3) , 687-690
- https://doi.org/10.1121/1.389853
Abstract
The effect of interposing a half-wavelength [acoustic] membrane between the transducer and the load in mechanically sectored ultrasonic scanners was investigated. The effects of impedance mismatch and detuning on broadband spectra were theoretically and experimentally studied. These studies indicate that for certain common cases, a bandwidth reduction of 20%-30% can be anticipated, resulting in a proportional decrease in the axial resolution.This publication has 0 references indexed in Scilit: