An Automatic Microcomputer System for Analysis of Monophasic Action Potentials
- 1 February 1990
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 13 (2) , 196-206
- https://doi.org/10.1111/j.1540-8159.1990.tb05070.x
Abstract
A computer system for rapid measurement and analysis of monophasic action potentials (MAPs) recorded in vivo was developed. MAPs recorded from the epicardium of mongrel dogs using a contact electrode were digitized by analog-to-digital conversion at a sampling rate of 1 kHz per channel for computer data acquisition. Activation time was detected using a sliding 10-point window at the location where the average positive dV/dT exceeds an adjustable threshold value in order to eliminate spurious detection due to baseline variability or motion artifact. Action potential duration (APD) was determined at 50% and 90% (APD50, APD90) repolarization levels at the first sample point below these detection levels. In addition, a tangent algorithm (APDtan) that detects peak negative dV/dT during repolarization was developed. APDtan was determined from the location of onset of activation to the intersection of tangent and baseline. APDtan allowed estimation of APD in the presence of subsequent premature beats when APD90 was not measurable. To validate activation time measurements, 4,600 action potentials were analyzed during fixed rate pacing. Over a range of paced coupling intervals from 200 to 1000 msec, an R2 value of 0.99968 and a slope of 0.9959 were obtained by linear regression between paced and calculated intervals. To validate APD measurements, 5035 action potentials were analyzed in five animals during fixed rate pacing (longer than 3 minutes) when action potential duration should be constant. Average coefficient of variation of 1.25%, 1.65%, and 1.14% were obtained for APD50, APD90, and APDtan, respectively. This algorithm provides a rapid and accurate method to analyze MAP activation and duration for basic physiological studies such as the determination of initiation of arrhythmias.Keywords
This publication has 25 references indexed in Scilit:
- Cycle length dependence of human action potential duration in vivo. Effects of single extrastimuli, sudden sustained rate acceleration and deceleration, and different steady-state frequencies.Journal of Clinical Investigation, 1988
- High resolution on-line analysis of cardiac action potentials using a motorola 68000 microcomputerJournal of Pharmacological Methods, 1988
- Relation of monophasic action potential recorded with contact electrode to underlying transmembrane action potential properties in isolated cardiac tissues: a systematic microelectrode validation studyCardiovascular Research, 1988
- Effects of pacing on triggered activity induced by early afterdepolarizations.Circulation, 1984
- A low-cost microcomputer system for automated analysis of intracellular cardiac action potentialsJournal of Pharmacological Methods, 1984
- Localization of regional myocardial ischemia by recording of monophasic action potentials.Circulation, 1984
- Intracardiac electrode detection of early ischaemia in man.Heart, 1983
- Long-term recording of monophasic action potentials from human endocardiumThe American Journal of Cardiology, 1983
- Cellular electrophysiologic characteristics of chronically infarcted myocardium in dogs susceptible to sustained ventricular tachyarrhythmiasJournal of the American College of Cardiology, 1983
- An algorithm for the machine calculation of complex Fourier seriesMathematics of Computation, 1965