Cantilever magnetometry in pulsed magnetic fields
- 1 November 1997
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 68 (11) , 4061-4065
- https://doi.org/10.1063/1.1148347
Abstract
The technique of cantilever magnetometry is shown to be functional in pulsed magnetic fields. Employing micromachined single crystalsilicon cantilevers and capacitance detection, we demonstrated a utilizable sensitivity to magnetic moment of 2.5×10 −12 Am 2 in magnetic fields to 36 T, representing an improvement of more than a factor of 10 over competing technologies. Torquemagnetization measurements on microcrystals of anisotropicsuperconductors are presented as evidence of the feasibility of the technique in long pulse magnets of pulse duration ∼0.1–1 s .Keywords
This publication has 6 references indexed in Scilit:
- Fast oscillations in (TMTSF)2XSynthetic Metals, 1997
- Seventy-two tesla non-destructive pulsed magnetic fields at AT&T Bell LaboratoriesPhysica B: Condensed Matter, 1994
- Today's and tomorrow's pulsed field facility in ToulousePhysica B: Condensed Matter, 1994
- Capacitance platform magnetometer for thin film and small crystal superconductor studiesAIP Conference Proceedings, 1992
- Non-destructive quasi-static pulsed magnetic fields at toulousePhysica B: Condensed Matter, 1989
- Magnetization Study of the Field-Induced Transitions in Tetramethyltetraselenafulvalenium Perchlorate,ClPhysical Review Letters, 1985