Delayed Time Response of the Continuous Cardiac Output Pulmonary Artery Catheter
- 1 December 1996
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 83 (6) , 1173-1177
- https://doi.org/10.1097/00000539-199612000-00007
Abstract
Previous studies of the accuracy of pulmonary artery catheters (PAC) which provide continuous cardiac output (CCO) monitoring have investigated the performance during steady-state conditions. We compared the response time to hemodynamic change using a CCO PAC and an ultrasonic flow probe (UFP). In five sheep, a CCO PAC was inserted, and an UFP for measurement of CCO was placed around the pulmonary artery via a left thoracotomy. Six interventions which rapidly alter cardiac output were studied: crystalloid bolus, balloon inflation in the inferior vena cava (IVC), IVC balloon deflation, dobutamine infusion, hemorrhage, and reinfusion of blood. Cardiac output measured before and after each intervention was used to calculate the total change caused by the intervention, and the time intervals from intervention to 20%, 50%, and 80% of that change were noted. For all interventions, the time response of CCO was significantly slower than UFP. The largest differences were seen with the rapid infusion of lactated Ringer's solution for which the time interval for 20% change was 7.3 +/- 2.3 min (mean +/- SD) for CCO versus 0.5 +/- 0.3 min for UFP. The time interval for 80% change was 14.5 +/- 4.1 min for CCO versus 1.8 +/- 0.9 min with UFP. The current study demonstrates clinically important time delays in the response of the CCO catheter. This delay must be considered when rapid alterations of the hemodynamic state may occur.Keywords
This publication has 22 references indexed in Scilit:
- Comparison of cardiac output measurement by continuous thermodilution with electromagnetometry in adult cardiac surgical patientsJournal of Cardiothoracic and Vascular Anesthesia, 1994
- Thoracic electrical bioimpedance measurement of cardiac output in postaortocoronary bypass patientsCritical Care Medicine, 1993
- COMPARISON OF THORACIC ELECTRICAL BIOIMPEDANCE AND THERMODILUTION FOR THE MEASUREMENT OF CARDIAC INDEX IN PATIENTS WITH SEVERE SEPSISBritish Journal of Anaesthesia, 1993
- Clinical Evaluation of Transtracheal Doppler for Continuous Cardiac Output EstimationAnesthesia & Analgesia, 1992
- Measurement of Pulmonary Blood Flow with Transesophageal Two-dimensional and Doppler EchocardiographyAnesthesiology, 1991
- Simultaneous Intraoperative Measurement of Cardiac Output by Thermodilution and Transtracheal DopplerAnesthesiology, 1991
- Intraoperative Estimation of Cardiac Output by Transesophageal Pulsed Doppler EchocardiographyAnesthesiology, 1991
- Continuous measurement of cardiac output with the use of stochastic system identification techniquesJournal of Clinical Monitoring and Computing, 1990
- Noninvasive Cardiac Output: Simultaneous Comparison of Two Different Methods with ThermodilutionAnesthesiology, 1990
- Simultaneous measurements of cardiac output by thermodilution, esophageal doppler, and electrical impedance in anesthetized patientsJournal of Cardiothoracic Anesthesia, 1988