Laser-Induced Fluorescence of Atoms in a Power-Modulated Inductively Coupled Plasma for Trace Detection and Diagnostic Study
- 1 August 1992
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 46 (8) , 1223-1225
- https://doi.org/10.1366/0003702924123908
Abstract
Laser-induced fluorescence of iron atoms during the plasma-off cycle of a power-modulated inductively coupled plasma has been performed. The low plasma background emission, and absence of radio-frequency interference, 1.4 ms after complete interruption of power to the ICP, results in an excellent environment for fluorescence detection. With the use of a simple nondispersive detection system, a limit of detection of 5 µg Fe/L was achieved. The iron atom population was found to remain constant for 1.2 ms after power interruption, indicating that the chemical environment during this period is inert. A spatial profile and ICP power study were conducted as diagnostic aids for optimization of the power-modulated ICP for both fluorescence and ionization detection.Keywords
This publication has 5 references indexed in Scilit:
- Laser-Induced Ionization of Atoms in a Power-Modulated Inductively Coupled PlasmaApplied Spectroscopy, 1992
- Spatially and temporally resolved emission and fluorescence in a power modulated inductively coupled plasmaSpectrochimica Acta Part B: Atomic Spectroscopy, 1990
- Fundamental Studies in the ICP Using a Sinusoidally Modulated Power InputApplied Spectroscopy, 1986
- Reduction of matrix ionization interference in laser-enhanced ionization spectrometryAnalytical Chemistry, 1981
- Atomic and ionic fluorescence spectrometry with pulsed dye laser excitation in the inductively-coupled plasmaAnalytica Chimica Acta, 1980