Assessment of a technique for 2D–3D registration of cerebral intra-arterial angiography
- 1 February 2004
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 77 (914) , 123-128
- https://doi.org/10.1259/bjr/27339681
Abstract
This study assesses the ability of a computer algorithm to perform automated 2D-3D registrations of digitally subtracted cerebral angiograms. The technique was tested on clinical studies of five patients with intracranial aneurysms. The automated procedure was compared against a gold standard manual registration, and achieved a mean registration accuracy of 1.3 mm (SD 0.6 mm). Two registration strategies were tested using coarse (1286128 pixel) or fine (2566256 pixel) images. The mean registration errors proved similar but registration of the lower resolution images was 3 times quicker (mean registration times 33 s, SD 13 s for low and 150 s SD 48 s for high resolution images). The automated techniques were considerably faster than manual registrations but achieved similar accuracy. The technique has several potential uses but is particularly applicable to endovascular treatment techniques.Keywords
This publication has 8 references indexed in Scilit:
- A Comparison of 2D-3D Intensity-Based Registration and Feature-Based Registration for NeurointerventionsPublished by Springer Nature ,2002
- 2D-3D Intensity Based Registration of DSA and MRA - A Comparison of Similarity MeasuresPublished by Springer Nature ,2002
- International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trialThe Lancet, 2002
- Intracranial Aneurysms: Clinical Value of 3D Digital Subtraction Angiography in the Therapeutic Decision and Endovascular TreatmentRadiology, 2001
- Three-dimensional Reconstructed Images after Rotational Angiography in the Evaluation of Intracranial Aneurysms: Surgical CorrelationNeurosurgery, 2000
- Analysis of cerebrovascular diseases by a new 3-dimensional computerised X-ray angiography systemNeuroradiology, 1998
- A comparison of similarity measures for use in 2-D-3-D medical image registrationIEEE Transactions on Medical Imaging, 1998
- Decomposition of transformation matrices for robot visionPattern Recognition Letters, 1984