Spectral Characteristics of a New Spectrometer Design for Atomic Emission Spectroscopy
- 1 August 1986
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 40 (6) , 813-821
- https://doi.org/10.1366/0003702864508241
Abstract
Recently a new commercial spectrometer has been developed that combines a linear photodiode array detector and an echelle spectrometer with a very unique dispersion/selection/recombination line selector. In this system, light from a source such as an ICP is first predispersed with the use of a low-resolution (0.5 M) concave grating polychromator. Desired narrow-wavelength regions are selected with the use of a slotted mask in the focal plane of this polychromator. The selected wavelengths transmitted by this mask are then recombined into a parallel quasi-white beam with the use of a concave mirror and a second grating. This collimated quasi-white beam is then directed to an echelle grating and the output spectrum from the echelle is finally focused on a linear photodiode array detector. The optical configuration and characteristics of this instrument will be presented along with spectra obtained with the use of an ICP source illustrating its spectral characteristics for atomic emission spectrometry.Keywords
This publication has 9 references indexed in Scilit:
- Application of pattern recognition and factor analysis to inductively coupled plasma optical emission spectraAnalytical Chemistry, 1986
- Signal-to-noise considerations for rapid scan atomic spectrometry with an inductively coupled plasmaSpectrochimica Acta Part B: Atomic Spectroscopy, 1985
- Some theoretical and practical considerations to the application of linear photodiode arrays for inductively coupled plasma emission spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 1985
- Cross-correlation masks for automatic detection of spectral informationSpectrochimica Acta Part B: Atomic Spectroscopy, 1981
- Signal-to-noise ratio performance characteristics of an inductively coupled plasmaAnalytical Chemistry, 1980
- Use of a Programmable Monochromator and a SIT Detector in Flame Atomic Emission SpectrometryBulletin of the Chemical Society of Japan, 1979
- Simultaneous Multielement Determinations by Atomic Absorption and Atomic Emission with a Computerized Echelle Spectrometer/Imaging Detector SystemPublished by American Chemical Society (ACS) ,1979
- Multiple entrance slit vidicon spectrometer for simultaneous multielement analysisAnalytical Chemistry, 1979
- Characteristics of Photodiode Arrays for Spectrochemical MeasurementsApplied Spectroscopy, 1976