Investigation of the Vaporization and Atomization of Boron in a Graphite Furnace Using Digital Imaging Techniques
- 1 July 1995
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 49 (7) , 1006-1016
- https://doi.org/10.1366/0003702953964804
Abstract
The distributions of atomic and molecular species of boron inside a graphite furnace were investigated with a charge-coupled device (CCD) camera. Extensive spatial nonuniformity of the distributions at high temperatures was observed for both species. Digital imaging by the CCD camera showed that molecular boron species are preferentially formed at the top of the graphite tube at 1800–2000°C and their concentrations increase toward the center of the tube. The lowest concentrations of molecular boron species occur at the furnace walls and especially at the bottom of the graphite furnace. The highest concentrations of boron atoms occur along the graphite tube walls, at the bottom. The absence of significant concentrations of molecular species desorbing from the initial site of analyte deposition suggests that the analyte boron species, which do not absorb light at 254 nm, are vaporized. Since the highest concentrations of the molecular boron species were detected away from the graphite tube wall, desorption of molecular boron species is not likely. A gas-phase thermal dissociation of boron oxide species, resulting in the formation of BO (g), is suggested at temperatures below the atomization temperature. Pre-atomization loss of boron as molecular boron species is clearly evident from the CCD images. The high appearance temperature (2300°C) and high concentrations of atomic boron at the graphite tube wall suggest that much of the atomic boron signal is probably the result of desorption of boron atoms from the decomposition of boron carbide (solid).Keywords
This publication has 22 references indexed in Scilit:
- Digital imaging of atomization processes in electrothermal atomizers for atomic absorption spectrometryAnalytical Chemistry, 1993
- Shadow spectral filming: a method of investigating electrothermal atomization. Part 3. Dynamics of longitudinal propagation of an analyte within graphite furnacesJournal of Analytical Atomic Spectrometry, 1993
- Spatially resolved spectroscopic measurements of CN, C2 and Al in graphite tube electrothermal atomizers under various operating conditionsJournal of Analytical Atomic Spectrometry, 1992
- Three-dimensional distributions of oxygen in graphite and metal tube atomizers for analytical atomic spectrometryJournal of Analytical Atomic Spectrometry, 1992
- Shadow spectral fiming: a method of investigating electrothermal atomization. Part 2. Dynamics of formation and structure of the absorption layer of aluminium, indium and gallium moleculesJournal of Analytical Atomic Spectrometry, 1992
- Shadow spectral filming: a method of investigating electrothermal atomization. Part 1. Dynamics of formation and structure of the absorption layer of thallium, indium, gallium and aluminium atomsJournal of Analytical Atomic Spectrometry, 1991
- The Effect of Internal Gas Flow on the Spatial Distribution of Sodium Atom within a Graphite Furnace AtomizerApplied Spectroscopy, 1990
- Spatial Mapping of Analyte Distribution within a Graphite Furnace AtomizerApplied Spectroscopy, 1988
- Time and spatial absorbance profiles within a graphite furnace atomizerSpectrochimica Acta Part B: Atomic Spectroscopy, 1982
- Optical system for gathering simultaneous time and spatial data from transient eventsAnalytical Chemistry, 1979