A New Ventriculoperitoneal Shunt with a Telemetric Intracranial Pressure Sensor: Clinical Experience in 94 Patients with Hydrocephalus
- 1 May 1997
- journal article
- Published by Wolters Kluwer Health in Neurosurgery
- Vol. 40 (5) , 931-935
- https://doi.org/10.1097/00006123-199705000-00009
Abstract
We have developed a telemetric intracranial pressure sensor (OSAKA telesensor; Nagano Keiki Seisakusyo Co. Ltd., Tokyo, Japan) and investigated the clinical usefulness of new ventriculoperitoneal shunting with an OSAKA telesensor, an on-off valve, and a programmable valve.The OSAKA telesensor was applied in 94 patients at Osaka Medical College between March 1986 and December 1995. Postoperative management, postoperative course, postural change of intracranial pressure (ICP), and zero drift of the OSAKA telesensor were investigated.ICP was measured easily in all patients with no inconvenience. The setting of the programmable valve was adjusted postoperatively 38 times in 25 patients. Postoperative ICP in the upright position was -14.9 +/- 4.5 mm Hg, and the difference in ICP between the supine and the upright position was about 20 mm Hg. The slope of the decrease in ICP with continuous postural changes from the supine to the upright position was steep until about 30 to 45 degrees but became incrementally less or even rose slightly about this angle. The initial slope was steeper when the shunt was open than when the shunt was closed; it was also steeper postoperatively than preoperatively. This initial slope, therefore, may indicate the intracranial compliance. Seven cases of shunt malfunction were diagnosed correctly by ICP measurements. Zero drift of the OSAKA telesensor was easily corrected by comparing the ICP reading from the telesensor with the puncture pressure in the on-off valve.The OSAKA telesensor is very useful in the postoperative care and pathophysiological evaluation of patients with hydrocephalus.Keywords
This publication has 11 references indexed in Scilit:
- Shunted hydrocephalus: Normal upright icp by csf gravity-flow control: A clinical study in young adultsSurgical Neurology, 1993
- Optimum Position for an Antisiphon Device in a Cerebrospinal Fluid Shunt SystemNeurosurgery, 1991
- Hydrocephalus: Overdrainage by ventricular shunts. A review and recommendationsSurgical Neurology, 1991
- Symptomatic low intracranial pressure in shunted hydrocephalusJournal of Neurosurgery, 1988
- Development of a New, Completely Implantable Intraventricular Pressure Meter and Preliminary Report of Its Clinical ExperienceNeurologia medico-chirurgica, 1981
- Body position and cerebrospinal fluid pressureJournal of Neurosurgery, 1976
- Body position and cerebrospinal fluid pressureJournal of Neurosurgery, 1976
- Chronic measurement of epidural pressure with an induction-powered oscillator transducerJournal of Neurosurgery, 1976
- Negative intracranial pressure hydrocephalus in adults with shunts and its relationship to the production of subdural hematomaJournal of Neurosurgery, 1974
- Effect of cerebrospinal fluid shunts on intracranial pressure and on cerebrospinal fluid dynamics: 2. A new technique of pressure measurements: results and concepts 3.A concept of hydrocephalusJournal of Neurology, Neurosurgery & Psychiatry, 1973