Transorbital Doppler diagnosis of intracranial arterial stenosis.
- 1 September 1986
- journal article
- research article
- Published by Wolters Kluwer Health in Stroke
- Vol. 17 (5) , 916-921
- https://doi.org/10.1161/01.str.17.5.916
Abstract
Forty-two intracranial internal carotid arteries were visualized by both arteriography and 2 MHz pulse ultrasonic Doppler examinations. The intracranial internal carotid artery Doppler signals were studied at 5-7 cm depth behind the eyelid using frequency spectral analysis. Stenosis criteria were developed and methods of avoiding confusion with collateral effects devised. Among 33 intracranial ICAs visualized on the arteriograms, 22 were normal without stenosis and 11 displayed some degree of stenosis ranging from 20 to 75 percent. An additional 7 were totally occluded. Doppler criteria of stenosis representing elevated frequencies and symmetrical prominent low frequencies (SPLF) were utilized and separated from collateral effects to provide an overall accuracy of the technique of 88 percent with a 95 percent specificity and a 73 percent sensitivity. The technique appears sufficiently promising to justify further development and utilization.This publication has 7 references indexed in Scilit:
- Methods to evaluate the sample volume of pulsed doppler systemsUltrasound in Medicine & Biology, 1984
- Musical murmurs in human cerebral arteries after subarachnoid hemorrhageJournal of Neurosurgery, 1984
- Determination of the Source of Musical Heart Murmur Using Two-Dimensional Doppler EchocardiographyPublished by Springer Nature ,1983
- Noninvasive transcranial Doppler ultrasound recording of flow velocity in basal cerebral arteriesJournal of Neurosurgery, 1982
- Vascular MurmursPublished by Springer Nature ,1981
- A 1-MHz Linear Phased Array Ultrasound System for Intracranial ImagingPublished by Springer Nature ,1980
- Local Doppler audio spectra in normal and stenosed carotid arteries in manUltrasound in Medicine & Biology, 1979