Arterial Stiffness as Underlying Mechanism of Disagreement Between an Oscillometric Blood Pressure Monitor and a Sphygmomanometer
- 1 October 2000
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Hypertension
- Vol. 36 (4) , 484-488
- https://doi.org/10.1161/01.hyp.36.4.484
Abstract
Oscillometric blood pressure devices tend to overestimate systolic blood pressure and underestimate diastolic blood pressure compared with sphygmomanometers. Recent studies indicate that discrepancies in performance between these devices may differ between healthy and diabetic subjects. Arterial stiffness in diabetics could be the underlying factor explaining these differences. We studied differences between a Dinamap oscillometric blood pressure monitor and a random-zero sphygmomanometer in relation to arterial stiffness in 1808 healthy elderly subjects. The study was conducted within the Rotterdam Study, a population-based cohort study of subjects aged 55 years and older. Systolic and diastolic blood pressure differences between a Dinamap and a random-zero sphygmomanometer were related to arterial stiffness, as measured by carotid-femoral pulse wave velocity. Increased arterial stiffness was associated with higher systolic and diastolic blood pressure readings by the Dinamap compared with the random-zero sphygmomanometer, independent of age, gender, and average mean blood pressure level of both devices. The beta-coefficient (95% CI) was 0.25 (0.00 to 0.50) mm Hg/(m/s) for the systolic blood pressure difference and 0.35 (0.20 to 0.50) mm Hg/(m/s) for the diastolic blood pressure difference. The results indicate that a Dinamap oscillometric blood pressure device, in comparison to a random-zero sphygmomanometer, overestimates systolic and diastolic blood pressure readings in subjects with stiff arteries.Keywords
This publication has 14 references indexed in Scilit:
- Determinants of the limits of agreement between the sphygmomanometer and the SpaceLabs 90207 device for blood pressure measurement in healthy volunteers and insulin-dependent diabetic patientsJournal Of Hypertension, 1998
- Does it matter whether blood pressure measurements are taken with subjects sitting or supine?Journal Of Hypertension, 1998
- Mechanisms by which the Hawksley random zero sphygmomanometer underestimates blood pressure and produces a non-random distribution of RZ valuesJournal of Human Hypertension, 1997
- Epidemiological Perspective of Body Position and Arm Level in Blood Pressure MeasurementBlood Pressure, 1994
- The Dinamap 1846SX automated blood pressure recorder: comparison with the Hawksley random zero sphygmomanometer under field conditions.Journal of Epidemiology and Community Health, 1992
- Arterial Wall Compliance in DiabetesDiabetic Medicine, 1992
- Determinants of disease and disability in the elderly: The Rotterdam elderly studyEuropean Journal of Epidemiology, 1991
- The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systemsJournal Of Hypertension, 1990
- Indirect Blood Pressure Measurements: A Matter of InterpretationAngiology, 1979
- The Principles Underlying Dinamap—a Microprocessor Based Instrument for the Automatic Determination of Mean Arterial PressureJournal of Medical Engineering & Technology, 1978