Dosimetric characteristics of Novalis shaped beam surgery unit
- 18 July 2002
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 29 (8) , 1729-1738
- https://doi.org/10.1118/1.1494830
Abstract
The dosimetric characteristics of a new dedicated radiosurgical treatment unit are systematically measured in terms of its percent depth dose, beam profile, and relative scatter factor. High-resolution diode detector, mini-ion-chamber detector, and conventional Kodak XV films are used to measure dosimetric data for a range of field sizes from 6x6 mm to 100x100 mm. The effects of collimator size, micro-multileaf collimator shape, and detector type on the dosimetric data are investigated. Results indicate that, with careful design, accurate dosimetric data could be acquired using either a dedicated diode detector or a mini-ion-chamber detector, and film detector. Special attention is required when measuring dosimetric data for small field sizes such as 6x6 mm.Keywords
This publication has 7 references indexed in Scilit:
- Total scatter factors and tissue maximum ratios for small radiosurgery fields: Comparison of diode detectors, a parallel‐plate ion chamber, and radiographic filmMedical Physics, 2000
- Assessment of new small-field detectors against standard-field detectors for practical stereotactic beam data acquisitionPhysics in Medicine & Biology, 1999
- Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgeryRadiotherapy and Oncology, 1999
- Physical characteristics of a miniature multileaf collimatorMedical Physics, 1999
- Comparison of Miniature Multileaf Collimation (MMLC) with circular collimation for stereotactic treatmentInternational Journal of Radiation Oncology*Biology*Physics, 1997
- Dosimetric characteristics of a commercial multileaf collimatorMedical Physics, 1995
- Doses on the central axes of narrow 6‐MV x‐ray beamsMedical Physics, 1990