Closed-circuit metabolic system with multiple applications
- 1 October 1986
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 61 (4) , 1582-1585
- https://doi.org/10.1152/jappl.1986.61.4.1582
Abstract
A closed-circuit metabolic system has been designed and tested for multiple applications. Air pressure within a closed chamber is regulated electronically while allowing for respiratory gas exchange. Compared with a previously reported standard indirect calorimetry system, the new device had by virtue of longer duration of measurement improved precision (coefficient of variation 3% vs. 14%) during studies of O2 consumption both at room temperature and at 5 degrees C. In addition, a more physiological atmospheric environment is maintained. This system has also been utilized for simultaneously labeling groups of up to 20 weanling rats with 18O2 over a 2-day period and for exposure of rats to a hyperoxic (84% O2), normobaric environment for 4-day periods. Potential applications include maintenance of pressure (hypobaric through hyperbaric) and O2 (hypoxic through hyperoxic) controlled environments, exposure to toxic gases, study of diurnal variations in metabolic rate, measurement of metabolic expenditure with activity, and adaptation to other species including humans.This publication has 2 references indexed in Scilit:
- Glucose and fatty acid kinetics in fasted rats: Effects of previous protein intakeMetabolism, 1982
- An experimental chamber for long-term studies of chronic hypoxia in small animalsJournal of Applied Physiology, 1965