The influence of the learning effect on automated perimetry in patients with suspected glaucoma
- 1 October 1989
- journal article
- research article
- Published by Wiley in Acta Ophthalmologica
- Vol. 67 (5) , 537-545
- https://doi.org/10.1111/j.1755-3768.1989.tb04105.x
Abstract
The influence of the learning effect on the outcome of automated perimetry was studied as a function of eccentricity. The sample comprised 20 patients with suspected glaucoma who were all naive to automated perimetry. Visual field examination of the right eye follwed by that of the left eye was undertaken on each of three successive days and after a further interval of 12 days using a customised full-field program of the Humphrey Field Analyser 630 (stimulus size III). The program comprised 60 points out to an eccentricity of 60'' with an interstimulus grid of 12''. Global, central, peripheral, superior and inferior mean sensitivity each significantly increased (P .ltoreq. 0.01) from the first to the second right eye examinations adn from the third to the fourth left eye examinations (P .ltoreq. 0.01). The global shortterm fluctuation, central mean defect and number of stimulus presentations decreased from the first to the second right eye examination (P 0.01). The order of examination between eyes and the interval between examination sessions influences the response recorded by automated perimetry.Keywords
This publication has 18 references indexed in Scilit:
- Serial examination of the normal visual field using Octopus automated projection perimetry Evidence for a learning effectActa Ophthalmologica, 1987
- The Implications of the Results of Computerized Perimetry in Normals for the Statistical Evaluation of Glaucomatous Visual FieldsPublished by Springer Nature ,1987
- A package for the statistical analysis of visual fieldsPublished by Springer Nature ,1987
- Asymmetry and Variation in the Normal Hill of VisionArchives of Ophthalmology (1950), 1986
- Changes in differential threshold in patients with glaucoma during prolonged perimetry.British Journal of Ophthalmology, 1983
- Deterioration of Threshold in Glaucoma Patients During PerimetryDocumenta Ophthalmologica Proceedings Series, 1983
- Changes of Glaucomatous Field Defects Analysis of OCTOPUS fields with programme DeltaPublished by Springer Nature ,1981
- TIME CHANGES OF CONTRAST THRESHOLDS DURING AUTOMATIC PERIMETRYActa Ophthalmologica, 1977
- STATIC PERIMETRY: ACCURACY AND FLUCTUATIONSActa Ophthalmologica, 1976