Effects of pressure and volume of the receiving chamber on the spatial distribution of regurgitant jets as imaged by color Doppler flow mapping. An in vitro study.
- 1 February 1991
- journal article
- abstracts
- Published by Wolters Kluwer Health in Circulation
- Vol. 83 (2) , 605-613
- https://doi.org/10.1161/01.cir.83.2.605
Abstract
Regurgitant jet dimensions imaged by color Doppler flow mapping have been used to evaluate the severity of valvular insufficiency in clinical studies. To study the effect of pressure and volume within the receiving chamber on the magnitude of spatial distribution of regurgitant jets assessed by color Doppler techniques, we designed a simple constant-flow model in which a jet was driven through a known orifice (1.5 mm2) into a compliant receiving chamber by a steady-flow pump. A distal tube at the outflow closed the system and maintained the volume of the chamber constant during pump operation. We varied flow rate from 60 to 270 ml/min into elastic balloons with different static compliances of 1, 2, 4.5, and 9 ml/mm Hg (pressures of 57, 28, 18, and 8 mm Hg, respectively); the balloons served as receiving chambers at the constant volume of 150 ml. We also evaluated the effect of different volumes of a receiving chamber (110, 130, and 150 ml and pressures of 5, 15, and 24 mm Hg) with a static compliance of 2 ml/mm Hg over the same range of flow rates. For each of the different balloons, jet area correlated linearly with the jet velocity across the orifice (r = 0.98, 0.99, 0.98, and 0.97) and also with flow rate (r = 0.97, 0.99, 0.98, and 0.99). At the same flow rate and volume of receiving chamber, however, the jet area imaged by color Doppler decreased as the pressure in the receiving chamber increased, although receiving-chamber volume was constant.(ABSTRACT TRUNCATED AT 250 WORDS)Keywords
This publication has 16 references indexed in Scilit:
- Review of hydrodynamic principles for the cardiologist: Applications to the study of blood flow and jets by imaging techniquesPublished by Elsevier ,2004
- Instrumentation and physical factors related to visualization of stenotic and regurgitant jets by Doppler color flow mappingJournal of the American College of Cardiology, 1988
- Computer analysis of Doppler color flow mapping images for quantitative assessment of in vitro fluid JetsJournal of the American College of Cardiology, 1988
- Color doppler echocardiographic assessment of the change in the mitral regurgitant volumeAmerican Heart Journal, 1987
- Clinical applications of a new type of real-time two-dimensional doppler flow imaging systemThe American Journal of Cardiology, 1984
- The development of real-time two-dimensional doppler echocardiography and its clinical significance in acquired valvular diseases With special reference to the evaluation of valvular regurgitation.Japanese Heart Journal, 1984
- Noninvasive diagnosis of aortic and mitral valve disease with pulsed-doppler spectral analysisThe American Journal of Cardiology, 1983
- Assessment of pulsed Doppler echocardiography in detection and quantification of aortic and mitral regurgitation.Heart, 1980
- Intracardiac flow pattern in mitral regurgitation studied with combined use of the ultrasonic pulsed doppler technique and cross-sectional echocardiographyThe American Journal of Cardiology, 1980
- The relationship of left atrial pressure and volume in patients with heart diseaseAmerican Heart Journal, 1964