A Simple Automated Colorimetric Method for Determination of N-Acetyl- β-d-Glucosaminidase
- 1 May 1982
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 19 (3) , 157-159
- https://doi.org/10.1177/000456328201900305
Abstract
An automated method for the determination of N-acetyl-β-d-glucosaminidase in serum or plasma using p-nitrophenol ( pNP) N-acetyl-β-d-glucosamine as substrate and a Pye Unicam AURA system is described. Normal samples had activities of 853 ± 146 (SD) nmol pNP liberated/ml/h, with intra-assay coefficient of variation 1·2% and inter-assay coefficient of variation 1·6%. Inhibition of enzyme activity by heparin in plasma samples can be reversed by the addition of calcium chloride to the buffer.This publication has 7 references indexed in Scilit:
- INCREASED PLASMA ACTIVITIES OF N-ACETYL-β-D-GLUCOSAMINIDASE ISOENZYMES IN HUMAN DIABETES MELLITUSClinical Endocrinology, 1979
- levels and diabetic microangiopathyClinica Chimica Acta; International Journal of Clinical Chemistry, 1979
- β-Glycosidases and Diabetic Microangiopathy II. An Insulin-dependent Isozyme of β-N AcetylglucosaminidaseThe Journal of Biochemistry, 1976
- Serum N-acetyl-beta-glucosaminidase Activity in Diabetic PatientsDiabetes, 1972
- Tay-Sachs DiseaseNew England Journal of Medicine, 1970
- SERUM β-GLUCURONIDASE, GLUCOSE TOLERANCE, AND ATHEROSCLEROTIC DISEASEThe Lancet, 1965
- Studies on glucosaminidase. 4. The fluorimetric assay of N-acetyl-β-glucosaminidaseBiochemical Journal, 1961