Automation of Nanoscale Microcapillary Liquid Chromatography−Tandem Mass Spectrometry with a Vented Column
- 14 May 2002
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 74 (13) , 3076-3083
- https://doi.org/10.1021/ac025529o
Abstract
To fully automate the sample introduction step for nanoscale microcapillary liquid chromatography−tandem mass spectrometry (LC−MS/MS) analyses, 75 μm i.d. × 14 cm capillary columns were interfaced with a commercial autosampler instrument using a novel procedure which allowed dilute peptide samples to be transferred from the AS loop injector to the nanoscale column at flow rates up to 5 μL min-1. On-column enrichment and desalting was demonstrated for large sample volumes (>40 μL) by constructing a vent 2 cm after the entrance to the packed bed of 5-μm ODS-AQ modified silica. Salts and nonretained solutes were removed via the vent, which allowed for column washing independent of the continuation of the bed into the electrospray source. Separations of test peptide mixtures demonstrated 50-nL elution peak volumes with low- to subfemtomole detection levels. In addition, a highly complex peptide mixture (outer membrane preparation from Psuedemonas aeruginosa) was efficiently separated with more than 100 proteins identified from a single reversed-phase LC−MS/MS analysis. Finally, the vented column (V-column) was utilized for on-line separations in a multidimensional chromatography/tandem MS experiment where large numbers of strong cation exchange chromatography fractions from a trypsinized yeast lysate were desalted, concentrated, and analyzed in a completely automated fashion. The procedures for constructing and using a V-column require minimal changes in current methods and equipment for nano-LC−MS analyses using columns of 100-μm diameter and smaller.Keywords
This publication has 14 references indexed in Scilit:
- Analysis of Proteins and Proteomes by Mass SpectrometryAnnual Review of Biochemistry, 2001
- Mass Spectrometry in ProteomicsChemical Reviews, 2001
- Peer Reviewed: Calixarenes: Designer Ligands for Chemical SensorsPublished by American Chemical Society (ACS) ,2001
- Probability-based protein identification by searching sequence databases using mass spectrometry dataElectrophoresis, 1999
- Protein Identification at the Low Femtomole Level from Silver-Stained Gels Using a New Fritless Electrospray Interface for Liquid Chromatography–Microspray and Nanospray Mass SpectrometryAnalytical Biochemistry, 1998
- A Microcapillary Column Switching HPLC−Electrospray Ionization MS System for the Direct Identification of Peptides Presented by Major Histocompatibility Complex Class I MoleculesAnalytical Chemistry, 1998
- Application of micro-electrospray liquid chromatography techniques to FT-ICR MS to enable high-sensitivity biological analysisJournal of the American Society for Mass Spectrometry, 1998
- High throughput protein characterization by automated reverse‐phase chromatography/electrospray tandem mass spectrometryProtein Science, 1998
- Direct Analysis and Identification of Proteins in Mixtures by LC/MS/MS and Database Searching at the Low-Femtomole LevelAnalytical Chemistry, 1997
- A Microscale Electrospray Interface for Online, Capillary Liquid Chromatography/Tandem Mass Spectrometry of Complex Peptide MixturesAnalytical Chemistry, 1995