Color Duplex Ultrasonography Is Insensitive for the Detection of Endoleak After Aortic Endografting: A Systematic Review
- 1 June 2005
- journal article
- research article
- Published by International Society of Endovascular Specialists in Journal of Endovascular Therapy
- Vol. 12 (3) , 297-305
- https://doi.org/10.1583/04-1479r.1
Abstract
To synthesize the available evidence regarding the diagnostic accuracy of color duplex ultrasonography (CDU) versus the accepted gold-standard of contrast-enhanced computed tomography (CT) for the detection and classification of endoleaks after aortic endografting. A systematic search of the literature was conducted using electronic bibliographical databases and other means to gather articles published between 1991 and 2004. Articles were scrutinized against inclusion/exclusion criteria that broadly followed the QUA-DAS quality assessment guidelines. The results of diagnostic CDU were expressed for each study as a 2times2 contingency table, and summary statistics (sensitivity/specificity with 95% confidence intervals [CI]) were calculated. Pooled and random effects meta-analyses were performed. Eight published studies and 2 unpublished studies from Charing Cross and St. George's Hospitals (711 patients, 1355 paired scans performed ≥1 month after endografting) were eligible for inclusion. From meta-analyses, the pooled sensitivity of CDU (versus CT as the gold standard) was 69% (95% CI 52% to 87%) and the specificity of CDU was 91% (95% CI 87% to 95%). These parameters did not appear to vary over time when a smaller dataset of 117 patients with 239 paired scans was used to compare CT and CDU specifically at 3, 12, and 24 months after endografting. Endoleak classification data, which was derived from only 5 small studies, indicated that CDU appeared to have better diagnostic accuracy in detecting type I or type III endoleaks compared with type II endoleaks; however, the data were insufficient for statistical analysis. CDU currently does not have sufficient diagnostic accuracy for the detection of all endoleaks in routine clinical practice. The diagnostic accuracy of CDU may improve if type II endoleaks are ignored.Keywords
This publication has 26 references indexed in Scilit:
- Nature, frequency, and predictors of secondary procedures after endovascular repair of abdominal aortic aneurysmJournal of Vascular Surgery, 2003
- Efficacy of ultrasound scan contrast agents in the noninvasive follow-up of aortic stent graftsJournal of Vascular Surgery, 2003
- Evaluation of Abdominal Aortic Aneurysm After Endoluminal TreatmentAmerican Journal of Roentgenology, 2002
- Prospective Evaluation of Magnetic Resonance Imaging After Endovascular Treatment of Infrarenal Aortic AneurysmsEuropean Journal of Vascular and Endovascular Surgery, 2001
- Systematic reviews in health care: Systematic reviews of evaluations of diagnostic and screening testsBMJ, 2001
- Role of Duplex Scan in Endoleak Detection After Endoluminal Abdominal Aortic Aneurysm RepairEuropean Journal of Vascular and Endovascular Surgery, 2000
- Realistic Expectations for Patients with Stent-graft Treatment of Abdominal Aortic Aneurysms. Results of a European Multicentre RegistryEuropean Journal of Vascular and Endovascular Surgery, 1999
- Endoleak after aortic stent graft repair: Diagnosis by color duplex ultrasound scan versus computed tomography scanJournal of Vascular Surgery, 1998
- Comparison of computed tomography and duplex imaging in assessing aortic morphology following endovascular aneurysm repairBritish Journal of Surgery, 1998
- Meta-analysis in clinical trialsControlled Clinical Trials, 1986