Response of a Large NaI(Tl) Crystal to High Energy Electrons
- 1 April 1963
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 34 (4) , 362-365
- https://doi.org/10.1063/1.1718365
Abstract
The resolution function of a 9‐in.‐diam by 10‐in.‐long cylindrical NaI(Tl) crystal for incident electrons of 55‐, 72‐, and 83‐MeV energy has been measured. A wedge magnet defined the electron energy. Spark chambers were used on either side of the magnet to insure against spurious scattering of electrons and to estimate the magnet system momentum spread. The most probable pulse height from the crystal varied linearly with electron energy. The pulse‐height distribution had a full width at half‐maximum of 13.4 +0.3 or −0.7%.Keywords
This publication has 2 references indexed in Scilit:
- Total-Absorption Scintillation Counter for High-Energy PhotonsReview of Scientific Instruments, 1962
- Response of a sodium-iodide scintillation spectrometer to 10- to 20-million-electron-volt electrons and x-raysJournal of Research of the National Bureau of Standards, 1956