An Implantable Neurocybernetic Prosthesis System
- 1 June 1990
- Vol. 31 (s2) , S33-S37
- https://doi.org/10.1111/j.1528-1157.1990.tb05846.x
Abstract
The neurocybernetic prosthesis (Cyberonics, Inc.) is an implantable, multiprogrammable pulse generator that delivers constant current electrical signals to the vagus nerve for the purpose of reducing the frequency and/or severity of epileptic seizures. The device is implanted in a subcutaneous chest pocket just below the clavicle, similar to cardiac pacemaker placement. The stimulation signal is transmitted from the prosthesis to the vagus nerve through a stimulation lead. The prosthesis can be programmed using any IBM-compatible personal computer with programming software and a programming wand. The electrodes used in the first group of patients were found to break at an unacceptable rate. Design modifications appear to have resolved this problem.Keywords
This publication has 6 references indexed in Scilit:
- Surgical Technique for Implantation of the Neurocybernetic ProsthesisEpilepsia, 1990
- Effects of Vagal Stimulation on Experimentally Induced Seizures in RatsEpilepsia, 1990
- Efficacy and Safety of Vagus Nerve Stimulation in Patients With Complex Partial SeizuresEpilepsia, 1990
- Prevention of Intractable Partial Seizures by Intermittent Vagal Stimulation in Humans: Preliminary ResultsEpilepsia, 1990
- Considerations for Safety with Chronically Implanted Nerve ElectrodesEpilepsia, 1990
- Cortical and subcortical patterns of response to afferent vagal stimulationExperimental Neurology, 1966