Output factors for blocked fields
- 1 June 1992
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 65 (774) , 517-522
- https://doi.org/10.1259/0007-1285-65-774-517
Abstract
This study investigates the output from blocked treatment fields with 6 MV and 18 MV photons. Square fields, defined by collimators, were blocked down to smaller symmetrical square, rectangular and asymmetrical rectangular fields using blocks on a tray beneath the collimators. Up to 7.5% difference was found between the output measured for the blocked field and the output calculated using a collimator output factor for the surface field size. A method is described by which output factors for the blocked fields can be measured and used in practice. Three methods of approximating these output factors are also presented.Keywords