A Simple Automated Method for Measuring Pressure–Volume Curves during Mechanical Ventilation
- 1 January 1999
- journal article
- research article
- Published by American Thoracic Society in American Journal of Respiratory and Critical Care Medicine
- Vol. 159 (1) , 275-282
- https://doi.org/10.1164/ajrccm.159.1.9802082
Abstract
Measurement of respiratory compliance is advocated for assessing the severity of acute respiratory failure (ARF). Recently, the administration of an automated constant flow of 15 L/min was proposed as a method easier to implement at the bedside than supersyringe or inspiratory occlusions methods. However, pressure–volume (P-V) curves were shifted to the right because of the resistive properties of the respiratory system. The aim of this study was to compare the P-V curves obtained using two constant flows—3 and 9 L/min—during volume-controlled mechanical ventilation with those obtained with the supersyringe and the inspiratory occlusions methods. Fourteen paralyzed patients with ARF were studied. The supersyringe and the inspiratory occlusions methods were performed according to usual recommendations. The new automated method was performed during volume-controlled mechanical ventilation by setting the inspiratory:expiratory ratio at 80%, the respiratory frequency at 5 breaths/min, and the tidal volume at 500 or 1,500 ml. These peculiar ventilatory settings were equivalent to administering a constant flow of 3 or 9 L/min during a 9.6-s inspiration. Esophageal and airway pressures were recorded. P-V curves obtained by the 3-L/min constant-flow method were identical to those obtained by the reference methods, whereas the P-V curve obtained by the 9-L/min constant flow was slightly shifted to the right. The slopes of the P-V curves and the lower inflection points were not different between all methods, indicating that the resistive component induced by administering a constant flow equal to or less than 9 L/min is not of clinical relevance. Because the 3-L/min constant-flow method is not artifacted by the resistive properties of the respiratory system and does not require any other equipment than a ventilator, it is an easy-to-implement, inexpensive, safe, and reliable method for measuring the thoracopulmonary P-V curve at the bedside.Keywords
This publication has 20 references indexed in Scilit:
- Impact of Positive End-expiratory Pressure on Chest Wall and Lung Pressure–Volume Curve in Acute Respiratory FailureAmerican Journal of Respiratory and Critical Care Medicine, 1997
- Pressure-volume curves in acute respiratory failure: automated low flow inflation versus occlusion.American Journal of Respiratory and Critical Care Medicine, 1997
- Titration of tidal volume and induced hypercapnia in acute respiratory distress syndrome.American Journal of Respiratory and Critical Care Medicine, 1995
- Volume-pressure curve of the respiratory system predicts effects of PEEP in ARDS: "occlusion" versus "constant flow" technique.American Journal of Respiratory and Critical Care Medicine, 1994
- A method for studying the static volume-pressure curves of the respiratory system during mechanical ventilationJournal of Critical Care, 1989
- Causes of error of respiratory pressure-volume curves in paralyzed subjectsJournal of Applied Physiology, 1987
- Volume/pressure curve of total respiratory system in paralysed patients: artefacts and correction factorsIntensive Care Medicine, 1987
- Total Respiratory Pressure-Volume Curves in the Adult Respiratory Distress SyndromeChest, 1984
- A Pulse Method of Measuring Respiratory System Compliance in Ventilated PatientsChest, 1981
- Measurement of Intraesophageal PressureJournal of Applied Physiology, 1955