Hollow-Fiber Flow Field-Flow Fractionation of Synthetic Polymers in Organic Solvents
- 23 August 2001
- journal article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 73 (19) , 4736-4742
- https://doi.org/10.1021/ac010144b
Abstract
A modified polyacrylonitrile (PAN) hollow-fiber membrane from a commercial source has been applied as the separation channel in flow field-flow fractionation (FFF). With the PAN membrane fiber, the application range of flow FFF could be extended to synthetic polymers that are soluble in a variety of organic solvents. The PAN membrane was shown to be resistant to hydrophobic solvents, such as dichloromethane (DCM), tetrahydrofuran (THF), ethyl acetate, and methyl ethyl ketone (MEK), as was illustrated by the successful fractionation of different polymer standards in these solvents. The system performance was assessed using polystyrene (PS) standards with ethyl acetate as the solvent. For a 100 kDa PS standard, the average recoverywas 57%, but for standards with a molar mass of 400 kDa and higher, 100% recovery was obtained. A linear relationship between peak area and injected mass was found. The run-to-run and fiber-to-fiber repeatability was determined using 100- and 400 kDa PS standards. The repeatability appeared to be satisfactory, with relative standard deviations < 2% for the retention times and < 5% for the recoveries of the standards. Plate numbers for the 400 kDa standard on different fibers were in the order of 110. From measurements on the fractionation of ferritin aggregates, it is concluded that the instrumental band-broadening is negligible. For an accurate determination of diffusion coefficients and molecular sizes based on retention times, calibration of the channel with standards appeared to be necessary. However, it was shown that the FFF system could be coupled to a multiangle light scattering (MALS) detector, thus providing an alternative on-line method for calibration. Expressions for the maximum obtainable plate number per unit of time have been derived for a hollow-fiber flow FFF system. It is shown that an increase in the system performance can be expected from a scaling down of the fiber diameter.Keywords
This publication has 16 references indexed in Scilit:
- Light scattering and the absolute characterization of macromoleculesPublished by Elsevier ,2001
- MEMBRANE PREPARATION | Hollow-Fibre MembranesPublished by Elsevier ,2000
- Micro‐FFF: Theoretical and practical aspects of reducing the dimensions of field‐flow fractionation channelsJournal of Microcolumn Separations, 1993
- Perturbations of the Retention Parameter Due to Sample Overloading in Hollow-Fiber Flow Field-Flow FractionationSeparation Science and Technology, 1993
- Properties of Hollow Fibers Used for Flow Field-Flow FractionationSeparation Science and Technology, 1993
- Effects of ionic strength of eluent on retention behavior and on the peak broadening process in hollow fiber flow field-flow fractionationJournal of Microcolumn Separations, 1991
- Zone broadening and dilution in rectangular and trapezoidal asymmetrical flow field-flow fractionation channelsAnalytical Chemistry, 1991
- Sample overloading effects in polymer characterization by field‐flow fractionationJournal of Applied Polymer Science, 1988
- Diffusion of polystyrene in tetrahydrofuranPolymer, 1977
- Flow-Field-Flow Fractionation: A Versatile New Separation MethodScience, 1976