Errors and limitations of the multimodality checking methods of defective spinal intrathecal pump systems. Case report
- 1 September 1993
- journal article
- case report
- Published by Springer Nature in Spinal Cord
- Vol. 31 (9) , 611-615
- https://doi.org/10.1038/sc.1993.97
Abstract
The loss of benefit from intrathecal baclofen, with increased spasticity and a discrepancy between the residual and the calculated volume content (underinfusion), made us suspect dysfunction of the intrathecal baclofen infusion in a paraplegic patient. Although all possible usual checking methods were used, no failure in the drug administration device (DAD) could be found. Despite some benefit from increasing the daily dose of baclofen, it has not been possible to control the spastic symptoms. Surgery was therefore decided upon, and a small leak at the end of the catheter tip was discovered. Surgeons and physicians should be aware that checking methods of DAD cannot exclude failure of the system. Their errors and limitations are discussed.Keywords
This publication has 8 references indexed in Scilit:
- Intrathecal baclofen for spasticity of spinal origin: seven years of experienceJournal of Neurosurgery, 1992
- Complications of intrathecal Baclofen deliveryBritish Journal Of Neurosurgery, 1992
- Chronic intrathecal baclofen administration for control of severe spasticityJournal of Neurosurgery, 1990
- Intrathecal baclofen for long-term treatment of spasticity: a multi-centre study.Journal of Neurology, Neurosurgery & Psychiatry, 1989
- Long-term intrathecal baclofen infusion for treatment of spasticityJournal of Neurosurgery, 1987
- Intrathecal Application of Baclofen in the Treatment of SpasticityPublished by Springer Nature ,1987
- CONTINUOUS INTRATHECAL BACLOFEN FOR SEVERE SPASTICITYThe Lancet, 1985
- INTRATHECAL BACLOFEN ALLEVIATES SPINAL CORD SPASTICITYThe Lancet, 1984