Pulse Height Response of Cd-Te to Heavy Ions
- 1 April 1981
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 28 (2) , 1546-1547
- https://doi.org/10.1109/tns.1981.4331463
Abstract
Pulse height spectra for alpha particles, fission fragments, and oxygen ions have been obtained for a thin-window surface-barrier Cd-Te detector. Although the detector can exhibit a large pulse-height defect and poorer resolution than Si detectors, it is still usable in many applications. In addition, the high stopping power of Cd-Te makes it suitable for simultaneous detection of energetic light and heavy ions, as well as γ rays, with a simple compact room-temperature detector.Keywords
This publication has 5 references indexed in Scilit: