After discussion of the biologic basis of echographic tissue differentiation and of the required standardization of the ultrasonic diagnostic device by means of tissue models (citrated blood), a technical standard is presented. This solid tissue model, which can be used an unlimited number of times, enables the investigator to determine or check the tissue sensitivity of his instrument at any time easily and quickly. The model consists of a synthetic resin (Wacker Silgel 504) to which a certain number of microbeads (glass beads S-100) are added before it hardens. The acoustic properties of this technical standard correspond to those of the biologic standard (citrated blood) including all important factors such as reflectivity, degree of scattering and sound attenuation.This publication has 6 references indexed in Scilit:
- Ten-year study on clinical echography in intraocular disease.1975
- A-scan echography and orbital disease.1975
- Ten-year study on clinical echography in orbital disease.1975
- [Additional experience with phantoms as an aid in ultrasonic diagnosis of intraocular tumors].1967
- Echographische Untersuchungen feingeweblicher StrukturenAlbrecht von Graefes Archiv für Ophthalmologie, 1967
- Weitere Erfahrungen mit Phantomen als Hilfsmittel zur Ultraschalldiagnostik intraocularer TumorenAlbrecht von Graefes Archiv für Ophthalmologie, 1967