Balloon angioplasty for treatment of in‐stent restenosis: feasibility, safety, and efficacy

Abstract
Sixty patients with 1 or 2 stainless steel intracoronary stents (Cook, Inc.) underwent balloon angioplasty for in-stent restenosis 1.5–13.5 months after stenting. Seventy-five in-stent redilatation procedures were performed. Seventy-three restenotic lesions (97%) were successfully recrossed and dilated, reducing the mean pre-angioplasty intrastent diameter stenosis from 77 ± 12% to 20 ± 11% residual. Although one angioplasty (1.3%) was complicated by non-Q-wave infarction, no angioplasty-related death, acute closure, need for additional stenting, emergent coronary bypass surgery, side branch occlusion, or vascular sequelae occurred. Post-procedure heparin was not used in 83% of successful cases. Most patients were discharged the day following redilatation (mean in-hospital stay 1.7 ± 1.3 days). At 5.4 ± 3.4 months following in-stent angioplasty, 84% of patients were in Canadian Cardiovascular Society class 0 or I. In conclusion, balloon dilatation in this stent for restenosis appears simple and efficacious in the short term, and may entail less risk than dilatation of unprotected coronary vessels. © 1994 Wiley-Liss,Inc..