Estimation of mean frequency and variance of ultrasonic Doppler signal by using second-order autoregressive model
- 1 May 1991
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 38 (3) , 172-182
- https://doi.org/10.1109/58.79600
Abstract
In order to estimate the mean frequency and variance of the diagnostic ultrasound Doppler signal in the presence of clutter noise, a new estimator using a second-order autoregressive (AR) model, called the AR estimator, is proposed. The sampled signal that contains information of both the Doppler signal and clutter is described by the second-order AR model with two poles. The mean frequency and variance of a unidirectional Doppler signal can be estimated, respectively, from the phase and the magnitude of the pole, with larger phase between the two poles. If the clutter is not completely rejected, all conventional estimators, including the autocorrelation (AC) estimator, result in erroneous estimations for the mean frequency and variance of the Doppler signal, whereas the AR estimator gives an accurate estimation. In the absence of clutter, however, the performance of both the AC and AR estimators are similar. If the blood flows in both directions in a sample volume and the clutter is rejected to the extent that it no longer obscures the Doppler signal, the proposed method can estimate simultaneously the mean frequencies and variances of both the forward and reverse blood flows. The performance of the proposed AR estimator was compared with that of the AC estimator by computer simulations and experiments, and it was found that when the number of available sampled data is small, the AR estimator does not require the use of a clutter filter, which simplifies Doppler signal detection.<>Keywords
This publication has 11 references indexed in Scilit:
- A new aliasing extension method for ultrasonic 2-dimensional pulsed Doppler systemsUltrasonic Imaging, 1989
- Bias and variance in the estimate of the Doppler frequency induced by a wall motion filterUltrasonic Imaging, 1989
- New multigate pulsed Doppler system using second-order samplingElectronics Letters, 1988
- Time-domain estimation of the center frequency and spread of Doppler spectra in diagnostic ultrasoundIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1988
- Simulation of Real-Time Frequency Estimators for Pulsed Doppler SystemsUltrasonic Imaging, 1986
- Color Doppler Flow ImagingPublished by Springer Nature ,1986
- A Comparison Between Mean Frequency Estimators for Multigated Doppler Systems with Serial Signal ProcessingIEEE Transactions on Biomedical Engineering, 1985
- Reduced-order autoregressive modeling for center-frequency estimationUltrasonic Imaging, 1985
- Power-spectrum centroid detection for doppler systems applicationsUltrasonic Imaging, 1980
- A Theoretical Study of the Scattering of Ultrasound from BloodIEEE Transactions on Biomedical Engineering, 1980