Testing and validating a homogeneous immunometric assay for interference

Abstract
We analyzed 95 sera, demonstrating interference in a previous study, with the Kryptor homogeneous time-resolved fluorescence resonance energy transfer carcinoembryonic antigen (CEA) immunoassay (Brahms AG, Berlin, Germany). Only one serum differed, i.e., 6.0 μg/l for Kryptor vs. 13.3 μg/l for a microtiter plate in-house immunofluorometric assay (IFMA), using both aggregated mouse immunoglobulins as blocker and capture monoclonal antibody (Mab) F(ab′)Kryptor-CEA assay results thus agreed with our inhouse CEA assay results, showing no interference. As the Kryptor-CEA assay antibodies were sensitive to interference and the Kryptor-CEA assay buffer did not reduce interference as efficiently as our in-house assay buffer, the Kryptor-CEA assay format was crucial for the absence of interference.