Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections
Top Cited Papers
Open Access
- 1 September 2011
- journal article
- research article
- Published by Springer Nature in Neuroradiology
- Vol. 54 (4) , 369-382
- https://doi.org/10.1007/s00234-011-0948-x
Abstract
Introduction The purpose of this study was to evaluate the safety and efficacy of the recently available flow diverter “pipeline embolization device” (PED) for the treatment of intracranial aneurysms and dissections. Methods Eighty-eight consecutive patients underwent an endovascular treatment of 101 intracranial aneurysms or dissections using the PED between September 2009 and January 2011. The targeted vessels include 79 (78%) in the anterior circulation and 22 (22%) in the posterior circulation. We treated 96 aneurysms and 5 vessel dissections. Multiple devices were implanted in 67 lesions (66%). Results One technical failure of the procedure was encountered. Immediate exclusion of the target lesion was not observed. Angiographic follow-up examinations were carried out in 80 patients (91%) with 90 lesions and revealed complete cure of the target lesion(s) in 47 (52%), morphological improvement in 32 lesions (36%), and no improvement in 11 lesions (12%). Six major complications were encountered: one fatal aneurysm rupture, one acute and one delayed PED thrombosis, and three hemorrhages in the dependent brain parenchyma. Conclusion Our experience reveals that the PED procedure is technically straightforward for the treatment of selected wide-necked saccular aneurysms, fusiform aneurysms, remnants of aneurysms, aneurysms with a high likelihood of failure with conventional endovascular techniques, and dissected vessels. While vessel reconstruction, performed after dissection, is achieved within days, remodeling of aneurysmal dilatations may take several months. Dual platelet inhibition is obligatory. Parenchymal bleeding into brain areas dependent on the target vessel is uncommon.Keywords
This publication has 29 references indexed in Scilit:
- A Comparison of Functional and Physical Properties of Self-Expanding Intracranial Stents [Neuroform3, Wingspan, Solitaire, Leo(+), Enterprise]min - Minimally Invasive Neurosurgery, 2011
- Association of Hemodynamic Characteristics and Cerebral Aneurysm RuptureAmerican Journal of Neuroradiology, 2010
- Flow-Diverter Stent for the Endovascular Treatment of Intracranial AneurysmsStroke, 2010
- Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective StudyPLOS ONE, 2010
- Very Late Thrombosis of a Pipeline Embolization Device ConstructOperative Neurosurgery, 2010
- High-Profile Flow Diverter (Silk) Implantation in the Basilar ArteryStroke, 2010
- Sole stenting of large and giant intracranial aneurysms with self-expanding intracranial stents—limits and complicationsActa Neurochirurgica, 2010
- Correlation Between Angiographic and Particle Image Velocimetry Quantifications of Flow Diverters in an In Vitro Model of Elastase-Induced Rabbit AneurysmsJournal of Biomechanical Engineering, 2009
- Endovascular Treatment of Small Intracranial Aneurysms: Three Alternatives to Coil Occlusionmin - Minimally Invasive Neurosurgery, 2006
- Increased Cell Opening and Prolapse of Struts of a Neuroform Stent in Curved Vasculature:Value of Angiographic Computed Tomography: Technical Case ReportOperative Neurosurgery, 2006