Pre-Clinical Testing of Hip Prosthetic Designs: A Comparison of Finite Element Calculations and Laboratory Tests
- 1 September 1993
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 207 (3) , 149-154
- https://doi.org/10.1243/pime_proc_1993_207_287_02
Abstract
To investigate the accuracy of finite element (FE) models for pre-clinical testing of unbonded hip prostheses, relative to aspects of load transfer and micromobility, two previously published laboratory experiments were simulated, using three-dimensional FE models. It was found for the load-transfer analyses that the experiment and the FE study revealed results that were very similar. The trends in the mobility experiments were also reproduced in the FE simulations, although quantitative differences were found. It is concluded that FE analysis can effectively be used for design evaluation of hip prostheses before prototypes are made.Keywords
This publication has 5 references indexed in Scilit:
- Rotational Movement of Femoral Components of Total Hip Replacements in Response to an Anteriorly Applied LoadEngineering in Medicine, 1988
- Femoral neck retention in hip arthroplasty A cadaver study of mechanical effectsActa Orthopaedica, 1988
- Finite-element-analysis and experimental investigation in a femur with hip endoprosthesisJournal of Biomechanics, 1983
- Three dimensional stress analysis of the human femurComputers in Biology and Medicine, 1977
- The elastic and ultimate properties of compact bone tissueJournal of Biomechanics, 1975