The Reliability of Doppler Ultrasound Techniques in the Assessment of Carotid Disease
- 1 June 1985
- journal article
- research article
- Published by SAGE Publications in Angiology
- Vol. 36 (6) , 333-340
- https://doi.org/10.1177/000331978503600601
Abstract
The reliability of continuous wave (CW) Doppler imaging and Duplex scan ning in the assessment of carotid artery disease has been evaluated prospectively in comparison with biplanar angiography. Of 130 comparisons the sensitivity of Doppler imaging was 89% and specificity 99% in the detection of ≥ 50% inter nal carotid stenosis, and 87% and 99% respectively in the diagnosis of internal carotid occlusion. Of 118 comparisons the sensitivity of Duplex scanning was 93% and specificity 98% in the detection of ≥ 50% internal carotid stenosis, and 93% and 99% respectively in the diagnosis of internal carotid occlusion. CW Doppler imaging and Duplex scanning are accurate techniques in the diagnosis and categorisation of internal carotid artery disease. CW Doppler imaging is an effective screening technique for the presence of ≥ 50% internal carotid artery stenosis. Duplex scanning also has the potential for improved detection of early internal carotid artery disease and study of the natural history of atherosclerosis.This publication has 12 references indexed in Scilit:
- The influence of ultrasonic duplex scanning on the management of carotid artery diseaseBritish Journal of Surgery, 1983
- Problems of carotid Doppler scanning which can be overcome by using frequency analysis.Stroke, 1982
- Correlation of peak Doppler frequency with lumen narrowing in carotid stenosis.Stroke, 1982
- Carotid artery disease: A prospective evaluation of pulsed doppler imagingUltrasound in Medicine & Biology, 1981
- Ultrasonic duplex scanning for disease of the carotid artery.Circulation, 1981
- Comparison of angiography with continuous wave Doppler ultrasound in the assessment of extracranial arterial diseaseJournal of Neurology, Neurosurgery & Psychiatry, 1981
- NON-INVASIVE STUDY OF THE CAROTID ARTERYThe Lancet, 1981
- Quantitation of carotid stenosis with continuous-wave (C-W) Doppler ultrasound.Stroke, 1979
- Color coded ultrasonic differential velocity arterial scanner (Echoflow)Ultrasound in Medicine & Biology, 1978
- Doppler cerebrovascular examination: improved results with refinements in technique.Stroke, 1977