Performance and reliability of five commercial enzyme-linked immunosorbent assay kits in screening for anti-human immunodeficiency virus antibody in high-risk subjects
- 1 August 1988
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 26 (8) , 1496-1500
- https://doi.org/10.1128/jcm.26.8.1496-1500.1988
Abstract
Anti-human immunodeficiency virus enzyme-linked immunosorbent assay kits marketed by Electro-Nucleonics Inc. (ENI), Genetic Systems Corp. (GSC), Organon Teknika Inc. (OTI), Ortho Diagnostic Systems Inc. (ODSI), and Wellcome Diagnostics (WD) were evaluated by using 289 randomly selected serum samples from a high-risk population and 52 serum samples likely to produce false-positive results. The radioimmunoprecipitation assay was used as the reference test. Sensitivities ranged from 96.51% (ODSI, WD) to 97.67% (ENI, GSC, OTI). Sera showing antibodies to viral glycoproteins only produced the false-negative results. Specificities ranged from 99.6%(ENI, GSC, ODSI, OTI) to 100% (WD). False-positive results were obtained with sera from patients with autoimmune disease or Epstein-Barr virus ifnection. Only results from GSC and OTI kits were distributed in two compact clusters well segregated on either side of the cutoff point. ODSI and GSC kits had the best intralot reproducibility. The GSC kit had the best interlot reproducibility. Cutoff values for ODSI and GSC kits were the least variable. Intraplate repeatability was good for all kits. Sample localization was not an important source of variability. Our results do not point out one outstanding kit among the five evaluated. However, the GSC kit showed the best overall results.This publication has 13 references indexed in Scilit:
- ANTIBODY RESPONSE IN PRIMARY HUMAN IMMUNODEFICIENCY VIRUS INFECTIONThe Lancet, 1987
- Screening and diagnostic performance of enzyme immunoassay for antibody to lymphadenopathy-associated virusJournal of Clinical Microbiology, 1987
- Patterns of Antibody Response in Individuals Infected with the Human Immunodeficiency VirusThe Journal of Infectious Diseases, 1987
- Comparison of indirect immunofluorescence and Western blot for detection of anti-human immunodeficiency virus antibodiesJournal of Clinical Microbiology, 1987
- Evaluation of a new assay for antibodies to LAV/HTLV IIIJournal of Virological Methods, 1986
- Predictive Value of a Screening Test for Antibodies to HTLV-IIIAmerican Journal of Clinical Pathology, 1986
- EVALUATION OF THE INDIRECT IMMUNOFLUORESCENCE ASSAY AS A CONFIRMATORY TEST FOR DETECTING ANTIBODIES TO THE HUMAN-IMMUNODEFICIENCY-VIRUS1986
- WHICH ANTI-HTLV III/LAV ASSAYS FOR SCREENING AND CONFIRMATORY TESTING?The Lancet, 1985
- Isolation of a T-Lymphotropic Retrovirus from a Patient at Risk for Acquired Immune Deficiency Syndrome (AIDS)Science, 1983
- Thermal gradients in microtitration plates. Effects on enzyme-linked immunoassayJournal of Immunological Methods, 1981