Miniature, high-resolution, quadrupole mass-spectrometer array
- 1 March 1997
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 68 (3) , 1393-1397
- https://doi.org/10.1063/1.1147947
Abstract
A miniature quadrupolemass spectrometer array consisting of 16 rods in a 4×4 array is reported. Each rod is 25 mm in length and 2 mm in diameter. The ionizer is of a miniature Nier-type design, and the detector is a channel-type electron multiplier. Operating frequencies are 5.3, 7.1, and 12.9 MHz. The mass range demonstrated herein is 1–300 u; and the resolution of the system is 0.1–0.5 u (full width at half-maximum), or m/Δm=600 . The present sensitivity is calculated and measured to be approximately 1×10 12 counts/Torr s.Keywords
This publication has 3 references indexed in Scilit:
- High-pressure effects in miniature arrays of quadrupole analyzers for residual gas analysis from 10−9 to 10−2 TorrJournal of Vacuum Science & Technology A, 1996
- Electromagnetic traps for charged and neutral particlesReviews of Modern Physics, 1990
- Das elektrische Massenfilter als Massenspektrometer und IsotopentrennerZeitschrift für Physik, 1958