Errors in Measuring Water Potentials of Small Samples Resulting from Water Adsorption by Thermocouple Psychrometer Chambers
- 1 September 1985
- journal article
- research article
- Published by Oxford University Press (OUP) in Plant Physiology
- Vol. 79 (1) , 184-188
- https://doi.org/10.1104/pp.79.1.184
Abstract
The adsorption of water by thermocouple psychrometer assemblies is known to cause errors in the determination of water potential. Experiments were conducted to evaluate the effect of sample size and psychrometer chamber volume on measured water potentials of leaf discs, leaf segments, and sodium chloride solutions. Reasonable agreement was found between soybean (Glycine max L. Merr.) leaf water potentials measured on 5-millimeter radius leaf discs and large leaf segments. Results indicated that while errors due to adsorption may be significant when using small volumes of tissue, if sufficient tissue is used the errors are negligible. Because of the relationship between water potential and volume in plant tissue, the errors due to adsorption were larger with turgid tissue. Large psychrometers which were sealed into the sample chamber with latex tubing appeared to adsorb more water than those sealed with flexible plastic tubing. Estimates are provided of the amounts of water adsorbed by two different psychrometer assemblies and the amount of tissue sufficient for accurate measurements of leaf water potential with these assemblies. It is also demonstrated that water adsorption problems may have generated low water potential values which in prior studies have been attributed to large cut surface area to volume ratios.This publication has 8 references indexed in Scilit:
- Ratio of Cut Surface Area to Leaf Sample Volume for Water Potential Measurements by Thermocouple PsychrometersPlant Physiology, 1984
- Field Use of Screen‐Caged Thermocouple Psychrometers in Sample Chambers1Crop Science, 1983
- Alterations in the Components of Peanut Leaf Water Potential during DesiccationJournal of Experimental Botany, 1981
- Water Content-Potential Relationship in Soya Bean: Changes in Component Potentials for Mature and Immature Leaves under Field Conditions*Annals of Botany, 1978
- Water Potential in Excised Leaf TissuePlant Physiology, 1978
- Water Potential Increase in Sliced Leaf Tissue as a Cause of Error in Vapor Phase Determinations of Water PotentialPlant Physiology, 1969
- A Vapour-pressure Instrument for the Measurement of Leaf and Soil Water PotentialJournal of Experimental Botany, 1965
- The Peltier Effect and its Use in the Measurement of Suction PressureJournal of Experimental Botany, 1951