Design, construction, performance, and use of a frequency-swept bridge detector for ion cyclotron resonance mass spectrometry are described. Special features include characterization and simple automatic correction of phase shift to allow broadband detection. The result is a detection system that may be used either at constant field or constant frequency. Drift-mode operation is simplified in that it may be satisfactorily used without the various signal modulation schemes used in previous detectors. In the trapped mode the detector may be pulsed to control the timing of ion detection. This detector makes it possible to do frequency-swept double resonance experiments which provide spectra of all the product ions of a given reactant ion. Circuit schematics and typical frequency- and field-swept spectra are shown.