Randoms variance reduction in 3D PET
- 1 January 1999
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 44 (4) , 941-954
- https://doi.org/10.1088/0031-9155/44/4/010
Abstract
In positron emission tomography (PET), random coincidence events must be removed from the measured signal in order to obtain quantitatively accurate data. The most widely implemented technique for estimating the number of random coincidences on a particular line of response is the delayed coincidence channel method. Estimates obtained in this way are subject to Poisson noise, which then propagates into the final image when the estimates are subtracted from the prompt signal. However, this noise may be reduced if variance reduction techniques similar to those used in normalization of PET detectors are applied to the randoms estimates prior to use. We have investigated the effects of randoms variance reduction on noise-equivalent count (NEC) rates on a whole-body PET camera operating in 3D mode. NEC rates were calculated using a range of phantoms representative of situations that might be encountered clinically. We have also investigated the properties of three randoms variance reduction methods (based on algorithms previously used for normalization) in terms of their systematic accuracy and their variance reduction efficacy, both in phantom studies and in vivo. Those algorithms investigated that do not make assumptions about the spatial distribution of random coincidences give the best estimates of the randoms distribution. With the camera used, which has a limited axial extent (10.8 cm) and a large ring diameter (102 cm), the gains in image signal-to-noise ratio obtained with this technique ranged from to , depending on object size, activity distribution and the amount of activity in the field of view. Larger gains would be expected if this technique were to be employed on cameras of greater axial extent and smaller ring diameter.Keywords
This publication has 11 references indexed in Scilit:
- Experience with fully 3D PET and implications for future high-resolution 3D tomographsPhysics in Medicine & Biology, 1998
- The effect of activity outside the direct field of view in a 3D-only whole-body positron tomographPhysics in Medicine & Biology, 1998
- Algorithms for calculating detector efficiency normalization coefficients for true coincidences in 3D PETPhysics in Medicine & Biology, 1998
- Optimization of noise-equivalent count rates in 3D PETPhysics in Medicine & Biology, 1996
- Model-based scatter correction for fully 3D PETPhysics in Medicine & Biology, 1996
- Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents countsIEEE Transactions on Nuclear Science, 1990
- PET system calibrations and corrections for quantitative and spatially accurate imagesIEEE Transactions on Nuclear Science, 1989
- Analytic 3D image reconstruction using all detected eventsIEEE Transactions on Nuclear Science, 1989
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1986
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1981