Design and characterization of a dynamic multileaf collimator
- 1 October 1998
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 43 (10) , 3149-3155
- https://doi.org/10.1088/0031-9155/43/10/033
Abstract
The characteristics of a prototype computer-assisted dynamic multileaf collimator (DMLC), specifically designed for small-field conformal radiotherapy, were evaluated at the Istituto Nazionale Tumori of Milan. The collimating device consists of two opposing banks of 16 pairs of 8 cm thick, 3.6 mm wide tungsten leaves and allows shaping of a radiation field up to a size of at the isocentre. The screening thickness of each leaf is 6.25 mm at the accelerator gantry isocentre. The leaves have a trapezoidal cross section and move along an arched path, thus providing a `double focused' collimation system. The DMLC was installed on the head of a Varian Clinac 2100C linear accelerator. Mechanical and dosimetric evaluations were performed to test the stability of the mechanical isocentre and to determine leaf leakage, penumbra width, accuracy of leaf positions and uniformity of leaf speed. Displacement of the mechanical isocentre was less than 1 mm at all gantry angles. Standard radiographic films exposed to 6 MV x-ray radiation were used for dosimetric evaluations. Leakage between leaves was less than 2.5%, and leakage through abutted leaves was less than 5.5%. The penumbra width between 20% and 80% isodose at different positions of leaf banks was 2.7 mm in the direction of the leaf motion and 3.1 mm along the side of the leaf with a standard deviation of 0.2 mm in both directions. Accuracy in the positioning of the leaf was 0.3 mm, whereas the maximum repositioning error was less than 0.2 mm. Finally, during movement of the leaves at the maximum speed of , the standard deviation of the leaf positioning error was 0.2 mm, proving an accurate uniformity of leaf speed.Keywords
This publication has 11 references indexed in Scilit:
- Dosimetric comparison of an integrated multileaf-collimator versus a conventional collimatorPhysics in Medicine & Biology, 1997
- Clinical implementation of a commercial multileaf collimator: Dosimetry, networking, simulation, and quality assuranceInternational Journal of Radiation Oncology*Biology*Physics, 1995
- The design and performance characteristics of a multileaf collimatorPhysics in Medicine & Biology, 1994
- Static field conformal stereotactic radiosurgery: Physical techniquesInternational Journal of Radiation Oncology*Biology*Physics, 1994
- Characterization of a multileaf collimator systemInternational Journal of Radiation Oncology*Biology*Physics, 1993
- Computer systems and mechanical tools for stereotactically guided conformation therapy with linear acceleratorsInternational Journal of Radiation Oncology*Biology*Physics, 1992
- Evaluation of multileaf collimator design for a photon beamInternational Journal of Radiation Oncology*Biology*Physics, 1992
- Variables associated with the development of complications from radiosurgery of intracranial tumorsInternational Journal of Radiation Oncology*Biology*Physics, 1991
- Comparison of the penumbra between focused and nondivergent blocks—Implications for multileaf collimatorsMedical Physics, 1991
- Improved linac dose distributions for radiosurgery with irregularly shaped fieldsInternational Journal of Radiation Oncology*Biology*Physics, 1990