A Mueller bridge set for cryogenic temperature measurements
- 1 August 1966
- journal article
- research article
- Published by IOP Publishing in Journal of Scientific Instruments
- Vol. 43 (8) , 576-580
- https://doi.org/10.1088/0950-7671/43/8/329
Abstract
An a. c. [alternate current] Mueller bridge set for resistance thermometry at cryogenic temperature is described. A commercial tuned null detector is used at an operating frequency of 1025 c/s. The set includes a high stability oscillator, line reject filter, phase shifter, Q multiplier and selector box. The latter permits the dissipation in the thermometers not being measured to be maintained at the operating level with direct current. A temperature change of the order of 10 udegk can be detected with 10-8 w applied to the thermometer.Keywords
This publication has 3 references indexed in Scilit:
- Transport Properties of Cobalt at Low TemperaturesPhysica Status Solidi (b), 1965
- Resistance Thermometer Bridge for Measurement of Temperatures in the Liquid Helium RangeReview of Scientific Instruments, 1958
- The Low Temperature Characteristics of Carbon-Composition ThermometersReview of Scientific Instruments, 1952