Dynamic exercise imaging with an MR‐compatible stationary cycle within the general electric open magnet
Open Access
- 18 March 2003
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 49 (3) , 581-585
- https://doi.org/10.1002/mrm.10364
Abstract
Many cases of muscular ischemia do not manifest without increased metabolic demand. Hence, diagnosis of intermittent claudication often requires inducing physiologic challenge, such as by exercise. Cine phase-contrast MRI can concurrently acquire cross-sectional vascular anatomy and through-plane blood velocities, enabling blood flow rate quantification. An MR-compatible stationary cycle was designed, constructed, and tested for flow quantification in large arteries during lower-limb exercise in a General Electric Signa SP 0.5 T open magnet. The cycle demonstrated smooth cycling during image acquisition, has freewheeling capability, is adjustable for subject size and strength, and can quantify workload. A healthy 59-year-old male was imaged at the supraceliac and infrarenal levels of the abdominal aorta at rest and during exercise. An exercise workload of 47.9 W was achieved. His heart rate increased from 52 to 78 bpm, supraceliac flow increased from 1.7 to 3.7 L/min, and infrarenal flow increased from 0.4 to 3.2 L/min from rest to exercise. Magn Reson Med 49:581–585, 2003.Keywords
This publication has 15 references indexed in Scilit:
- Quantification of Wall Shear Stress in Large Blood Vessels Using Lagrangian Interpolation Functions with Cine Phase-Contrast Magnetic Resonance ImagingAnnals of Biomedical Engineering, 2002
- In Vivo Quantification of Blood Flow and Wall Shear Stress in the Human Abdominal Aorta During Lower Limb ExerciseAnnals of Biomedical Engineering, 2002
- Diagnosis and Treatment of Chronic Arterial Insufficiency of the Lower Extremities: A Critical ReviewCirculation, 1996
- Shear Stress Modulates Expression of Cu/Zn Superoxide Dismutase in Human Aortic Endothelial CellsCirculation Research, 1996
- MR perfusion imaging of the kidney pre- and post-dipyridamole stressJournal of Magnetic Resonance Imaging, 1996
- Assessment of hemodynamic effects of angiotensin-converting enzyme inhibitor therapy in chronic aortic regurgitation by using velocity-encoded cine magnetic resonance imagingAmerican Heart Journal, 1996
- Simultaneous temporal resolution of cardiac and respiratory motion in MR imaging.Radiology, 1995
- Chronic exercise in dogs increases coronary vascular nitric oxide production and endothelial cell nitric oxide synthase gene expression.Circulation Research, 1994
- Chronic mesenteric ischemia: evaluation with phase-contrast cine MR imaging.Radiology, 1994
- Chronic exercise enhances endothelium-mediated dilation of epicardial coronary artery in conscious dogs.Circulation Research, 1993