Construction and Characteristics of Evaporated Nickel Bolometers
- 1 June 1947
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 18 (6) , 429-435
- https://doi.org/10.1063/1.1740966
Abstract
Details are given of the construction of an evaporated nickel bolometer. These bolometers are approximately 200A thick and are evaporated onto a nitrocellulose pellicle approximately 1000A thick. The pellicle is supported on a glass base in which is a sandblasted groove. This groove is immediately below the area of the pellicle which is covered with the bolometer strip. The techniques are described which were used to measure the bolometer characteristics. The sensitivity and frequency response of the bolometer are discussed. A typical bolometer has a time constant of 0.004 second, and a threshold of 3.3×10−8 watt for radiation modulated sinusoidally at 30 cycles per second where the band width of the system is 100 cycles.Keywords
This publication has 3 references indexed in Scilit:
- An Investigation of the Properties of Evaporated Metal Bolometers*Journal of the Optical Society of America, 1947
- High Transmission Windows for Radiation of 3 to 14 Microns Wave-LengthJournal of the Optical Society of America, 1947
- The Optical Properties of Metallic and Crystalline PowdersJournal of the Optical Society of America, 1933