Eye dose measurements using conventional and rare-earth screens during tomography of the para-nasal sinuses
- 1 April 1977
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 50 (592) , 266-270
- https://doi.org/10.1259/0007-1285-50-592-266
Abstract
Eye dose measurements have been performed when using medium speed conventional and rare-earth screen-film combinations during tomography of the para-nasal sinuses. The measurements show that using conventional intensifying screens with the A.P. view a total eye dose of about 20 rad may be given during an examination. This eye dose can be reduced by 98% using the P.A. position. If rare-earth screen/film combinations are employed the eye dose measured in the A.P. view was reduced by 75% of that obtained with conventional screens without detectable loss of image quality. A total eye dose reduction of about 99·5% was measured in the P.A. view with the rare-earth systems.This publication has 5 references indexed in Scilit:
- Film/Screen Considerations in TomographyRadiology, 1974
- THE FEASIBILITY OF THE PA PROJECTION FOR TOMOGRAPHY OF THE PETROUS BONEAmerican Journal of Roentgenology, 1974
- Noise Measurements On Rare-Earth Intensifying Screen SystemsPublished by SPIE-Intl Soc Optical Eng ,1974
- Patient Dose Distributions During Hypocycloidal TomographyRadiology, 1972
- Radiation Dose to Critical Organs During Petrous TomographyRadiology, 1970