Radiation dose reduction in the evaluation of scoliosis: an application of digital radiography.
- 1 October 1986
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 161 (1) , 175-181
- https://doi.org/10.1148/radiology.161.1.3763863
Abstract
This report documents the clinical testing of scanning beam digital radiography as an imaging method in patients with scoliosis. This type of digital imaging requires a skin exposure of only 2.4 mR (0.619 .mu.C/kg) per image, compared with the lowest possible posteroanterior screen-film exposure of 10 mR (2.58 .mu.C/kg) at the chest and 60 mR (15.48 .mu.C/kg) at the lumbar spine. Digital radiographic and screen-film images were obtained on multiple test objects and 273 patients. Scoliosis measurements using screen-film radiographs and digital radiographs were comparable to within a mean difference of 1.degree. at many different degrees of severity. The low-dose digital images were found to be useful and accurate for the detection and measurement of scoliosis after the first screen-film radiographs have excluded tumors and structural abnormalities.This publication has 9 references indexed in Scilit:
- Segmented-field radiography in scoliosisAmerican Journal of Roentgenology, 1985
- Fluoroscopy in children: low-exposure technology.Radiology, 1984
- A High Incidence of Spinal Curvature: A Study of 100 Young Female StudentsActa Orthopaedica, 1983
- Reduction of radiation exposure during radiography for scoliosis.Journal of Bone and Joint Surgery, 1982
- Normalized organ doses for various diagnostic radiologic proceduresAmerican Journal of Roentgenology, 1981
- The changing pattern of scoliosis treatment due to effective screening.Journal of Bone and Joint Surgery, 1981
- Risks of exposure to X-rays in patients undergoing long-term treatment for scoliosis.Journal of Bone and Joint Surgery, 1979
- ScoliosisJournal of Bone and Joint Surgery, 1978
- Low-Dosage X-Ray Imaging System Employing Flying Spot X-Ray Microbeam (Dynamic Scanner)Radiology, 1976