Abstract
Capillary array electrophoresis (CAE) is a novel technique, which allows for high throughput analysis of DNA fragments. When screening for mutations in whole populations or large patient groups it is necessary to have robust and well‐characterized setups for high throughput analysis. For large‐scale mutation screening, we have developed procedures for single strand conformation polymorphism (SSCP) assays using CAE (CAE‐SSCP) whereby we may increase both the sensitivity and the throughput compared to conventional SSCP analysis. In this study we have validated CAE‐SSCP by 1) comparing detection by slab‐gel based SSCP with CAE‐SSCP of mutations in the MYH7, MYL2, and MYL3 genes encoding sarcomere proteins from patients suffering from hypertrophic cardiomyopathy; and 2) by constructing a series of 185 mutants having substitution mutations, as well as insertion/deletion mutations, or some combinations of these, in different sequence contexts in four exons and different positions relative to the end of the amplicon (three from the KCNQ1 gene, encoding a cardiac potassium channel, and one from the TNNI3 gene encoding cardiac troponin I). The method identified 181 out of 185 mutations (98%), and the data suggest that the position of mutation in the fragment had no effect on the sensitivity. Analysis of the specificity of the method showed that only very few mutants could not be distinguished from each other and there were no false positives. Hum Mutat 21:116–122, 2003.Keywords
This publication has 23 references indexed in Scilit:
- Correlation of MFOLD-predicted DNA secondary structures with separation patterns obtained by capillary electrophoresis single-strand conformation polymorphism (CE-SSCP) analysisHuman Mutation, 2002
- Denaturing high-performance liquid chromatography: A reviewHuman Mutation, 2001
- Comparison of fluorescent single-strand conformation polymorphism analysis and denaturing high-performance liquid chromatography for detection of EXT1 and EXT2 mutations in hereditary multiple exostosesEuropean Journal of Human Genetics, 2000
- High-throughput single-strand conformation polymorphism analysis by automated capillary electrophoresis: Robust multiplex analysis and pattern-based identification of allelic variantsHuman Mutation, 1999
- High density synthetic oligonucleotide arraysNature Genetics, 1999
- Positional cloning reaches maturityCurrent Opinion in Genetics & Development, 1995
- Limitations in the use of SSCP analysisMutation Research - Fundamental and Molecular Mechanisms of Mutagenesis, 1993
- The Sensitivity of Single-Strand Conformation Polymorphism Analysis for the Detection of Single Base SubstitutionsGenomics, 1993
- How sensitive is PCR-SSCP?Human Mutation, 1993
- Rapid and sensitive detection of point mutations and DNA polymorphisms using the polymerase chain reactionGenomics, 1989