Antibody testing in Lyme disease. A comparison of results in four laboratories
- 22 December 1989
- journal article
- research article
- Published by American Medical Association (AMA) in JAMA
- Vol. 262 (24) , 3431-3434
- https://doi.org/10.1001/jama.262.24.3431
Abstract
To evaluate the interlaboratory and intralaboratory agreement in the performance of Lyme disease serological testing, we sent serum specimens from 132 outdoor workers in New Jersey to as many as four independent laboratories. These included one state department of health laboratory, one large commercial laboratory, and two research laboratories. The measurement of agreement employed, the kappa statistic, ranged from.45 to.53 among the four laboratories and from.50 to.54 within the commercial laboratory. These values represent low levels of agreement. The data suggest that Lyme disease serological testing procedures should be standardized so that Lyme disease test results are more comparable between laboratories. (JAMA. 1989;262:3431-3434)This publication has 4 references indexed in Scilit:
- ENZYME-LINKED IMMUNOSORBENT ASSAYS FOR THE DETECTION OF CLASS-SPECIFIC IMMUNOGLOBULINS TO BORRELIA BURGDORFERI1American Journal of Epidemiology, 1988
- Comparison of an indirect fluorescent-antibody test with an enzyme-linked immunosorbent assay for serological studies of Lyme diseaseJournal of Clinical Microbiology, 1984
- Antibody Response in Lyme Disease: Evaluation of Diagnostic TestsThe Journal of Infectious Diseases, 1984
- Enzyme-Linked Immunosorbent Assay and Indirect Immunofluorescence Assay for Lyme DiseaseThe Journal of Infectious Diseases, 1984