Design, fabrication, and evaluation of high frequency, single-element transducers incorporating different materials
- 7 August 2002
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 49 (2) , 169-176
- https://doi.org/10.1109/58.985701
Abstract
The performance of high frequency, single-element transducers depends greatly on the mechanical and electrical properties of the piezoelectric materials used. This study compares the design and performance of transducers incorporating different materials. The materials investigated include 1-3 lead zirconate titanate (PZT) fiber composite, lead titanate (PbTiO/sub 3/) ceramic, poly(vinylidene fluoride) (PVDF) film, and lithium niobate (LiNbO/sub 3/) single crystal. All transducers were constructed with a 3-mm aperture size and an f-number between 2 and 3. Backing and matching materials were selected based on design goals and fabrication limitations. A simplified coaxial cable tuning method was employed to match the transducer impedance to 50 /spl Omega/ for the PZT fiber composite and PbTiO/sub 3/ ceramic transducers. Transducers were tested for two-way loss and -6 dB bandwidth using the pulse/echo response from a flat quartz target. Two-way loss varied from 21 to 46 dB, and bandwidths measured were in the range from 47 to 118%. In vitro ultrasonic backscatter microscope (UBM) images of an excised human eye were obtained for each device and used to compare imaging performance. Both press-focusing and application of a lens proved to be useful beam focusing methods for high frequency. Under equal gain schemes, the LiNbO/sub 3/ and PbTiO/sub 3/ transducers provided better image contrast than the other materials.Keywords
This publication has 14 references indexed in Scilit:
- Fabrication of high frequency (25-75 MHz) single element ultrasonic transducersPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2003
- Transverse strain responses in electrostrictive poly(vinylidene fluoride-trifluoroethylene) films and development of a dilatometer for the measurementJournal of Applied Physics, 1999
- Passive materials for high-frequency ultrasound transducersPublished by SPIE-Intl Soc Optical Eng ,1999
- Design guidelines for medical ultrasonic arraysPublished by SPIE-Intl Soc Optical Eng ,1998
- Piezoceramics for high-frequency (20 to 100 MHz) single-element imaging transducersIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1997
- Ultrasonic transducers and arraysIEEE Engineering in Medicine and Biology Magazine, 1996
- A 40–100 MHz B-scan ultrasound backscatter microscope for skin imagingUltrasound in Medicine & Biology, 1995
- Fabrication of high frequency spherically shaped ceramic transducersIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1994
- The design and fabrication of high frequency poly(vinylidene fluoride) transducersUltrasonic Imaging, 1989
- The design of efficient broad-band piezoelectric transducersIEEE Transactions on Sonics and Ultrasonics, 1978