Absolute calibration of TFTR helium proportional countersa)
- 1 February 1995
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 66 (2) , 1247-1251
- https://doi.org/10.1063/1.1146015
Abstract
The TFTR helium proportional counters are located in the central five (5) channels of the TFTR multichannel neutron collimator. These detectors were absolutely calibrated using a 14 MeV neutron generator positioned at the horizontal midplane of the TFTR vacuum vessel. The neutron generator position was scanned in centimeter steps to determine the collimator aperture width to 14 MeV neutrons and the absolute sensitivity of each channel. Neutron profiles were measured for TFTR plasmas with time resolution between 5 and 50 ms depending upon count rates. The He detectors were used to measure the burnup of 1 MeV tritons in deuterium plasmas, the transport of tritium in trace tritium experiments, and the residual tritium levels in plasmas following 50:50 DT experiments.Keywords
This publication has 5 references indexed in Scilit:
- Absolute calibration of tokamak fusion test reactor neutron detectors for D-T plasma operationReview of Scientific Instruments, 1995
- Validation of spatial profile measurements of neutron emission in TFTR plasmas (invited)Review of Scientific Instruments, 1992
- Measurement of DT neutron emission from TFTR with helium-4 proportional recoil countersReview of Scientific Instruments, 1992
- Operation and cross calibration of the activation foil system on TFTRReview of Scientific Instruments, 1990
- TFTR multichannel neutron collimatorReview of Scientific Instruments, 1990