Pressure Regulator for a Low-Pressure Continuous-Flow Ionization Chamber

Abstract
A description is given of an automatic pressure‐regulating device for a continuous‐flow ionization chamber operated in the pressure range of 1 to 10 mm Hg. The pressure monitor consisted of an oil manometer, a light source, and two phototubes arranged whereby any small change in pressure gave rise to a corresponding electrical signal. This signal was used to control the power to a thermal gas flow valve. The regulator was capable of maintaining the pressure constant to within 0.03 mm Hg for a period of at least 24 hr.

This publication has 0 references indexed in Scilit: